TURA-X; Device for measuring the transformation ratio of three-phase true RMS transformers (TRMS) with integrated printer / Rapportimeter
TURA-X
- Ratio Measurement 0.8 to 50,000
- Test Voltages 1 V to 430 V
- High Accuracy 0.08%
- Built-in Printer
- TRUE 3-Phase
TURA-X are designed using advanced engineering technology to measure the turns ratio of single-phase and three-phase transformers.
Turns ratio measurement of phase-shifting transformers can be done by using TURA-X.
Its user-friendly software has fast, easy, and accurate turns ratio measurement features for current, voltage, and power transformers.
By ANSI/IEEE C57.12.90 measurement method, TURA-X can make precise results. TURA-X are one of the most accurate devices with its wide range ratio (0.8 to 50,000) measurement capability and high precision (0.08 %).
Besides the ratio measurement, TURA-X can also measure excitation current, phase angle, polarity and ratio error.
Even though TURA-X have a three-phase cable configuration, operators can make single-phase transformer tests as well.
TURA-X can detect vector groups automatically. TURA-X have wide operations range from high excitation value current transformers to high powered power transformers at substation with the 1V, 4V, 10V, 40 V, 100 V, 250 V and 430 V AC test voltages generation capability.
A 7-inch colour touch display allows TURA-X to show all measurement results on a single screen.
USB, Bluetooth (Option) and Flash Memory features allow TURA-X to control, record and store measurement results (up to 100 Test Record).
With the Data Management Platform (DMP Software), users can control TURA-X and analyse measurement results.
Operators can easily print the measurement results with the 2.28-inch built-in Printer of TURA-X.
Multi-language capability and user-friendly operation menu make it easy to control TURA-X, even by less trained staff.
TURA-X can control the tap changer (raise and lower) via its tap changer outputs.
TURA-X devices are light, compact and rugged with the protection class IP67 (case closed).
In addition to ratio tests, TURA-X can perform the Magnetic Balance Test as well.
Features
- Turns Ratio Measurement (Single-phase and TRUE three-phase measurement)
- Ranges from 0.8 to 50,000 ratio measurement
- High Accuracy (0.08 %)
- 1V, 4V, 10V, 40V, 100V, 250V & 430V AC test voltage
- Ratio Error Measurement
- Magnetic Balance Test
- Excitation Current, Phase Angle, Polarity Measurement
- Built-in Printer
- Optional Battery
- Automatic vector group detection feature
- Internal memory, USB Flash Drive
- PC control via USB cable
- Optional Bluetooth control and communication
Ordering Information
TURA-X; TRUE Three-Phase Transformer Turns Ratio Tester with Built-in Printer
TURA-X BLUE; TRUE Three-Phase Transformer Turns Ratio Tester with Built-in Bluetooth & Printer
TURA-XB; TRUE Three-Phase Transformer Turns Ratio Tester with Built-in Battery & Printer
TURA-XB BLUE; TRUE Three-Phase Transformer Turns Ratio Tester with Built-in Battery, Bluetooth & Printer
Measurement Parameters | Turn ratio measurement, Excitation current, Phase Angle, Polarity, Ratio Error(%), Vector Group Detection, Magnetic Balance | ||
Ratio Measurement Modes | CT Mode, PT Mode (Single-Phase and True Three-Phase) | ||
Measurement Method | ANSI / IEEE C57.12 | ||
Test Voltages | CT: 1 V e 4 V CT Mode (Phase-Neutral): 1, 4, 10, 40, 100, 250 V PT Mode (Phase-Phase): 1, 4, 10, 40, 100, 250 V, 430 V | ||
Ratio Range | 0,8 - 50000 | ||
Accuracy | Mode | Accuracy | Ratio |
CT Mode (1 V - 4 V) | 0,08% 0,1% | 0,8 - 399 400 - 4000 | |
PT Mode (10 V- 430 V) | 0,08% 0,15% 0,5% | 0,8 - 5000 5001 - 12000 12001 - 50000 | |
Phase angle measurement | 0 - 360 Degrees | ||
Excitation current | 2 A | ||
Power Supply | 100 - 240 V, 47/63 Hz | ||
Battery (Option) | Yes | ||
Memory | Up to 100 records (include up to 25 tap results) | ||
Printer | 2.28-inch printer | ||
Communication | USB 2.0/1.1 Standard-A, USB 2.0/1.1 Standard-B, Bluetooth (factory install Option) | ||
PC Software | DMP Software | ||
Display | 7-inch colour touch display | ||
Dimensions | 424 mm x 340 mm x 173 mm | ||
Weight | 7,1 kg | ||
Operating Temperature | -10 °C ... + 60 °C | ||
Protection Class | IP67 (closed housing) | ||
Set of Package | TURA-X, Power Cable, Ground Cable , 5 m (16.4 ft.) Standard Test Cable Set, 10 m (33 ft.) Extension Cable Set, Tap Changer Cable Set, USB Cable, Printer Paper (x2), USB flash drive, Instruction Manual (Soft Copy), DMP Software, Cable Bag | ||
Options | Hard Carrying Case, Battery, Bluetooth (factory install option) |